APPENDIX A - VFD CENTRIFUGE HMI SERVER SYSTEM
15 May 13 A-5
Appendix A-HMIServerSystem
5. Connect a CAT5 cable to Ethernet switch in Switchgear Box, and connect opposite end to
LAN 1 port on HMI Server (Figure A-6).
6. Tighten nuts properly on all cable glands to ensure proper sealing.
7. Connect a CAT5 cable from LAN 2 port of HMI Server to Ethernet switch port.
8. Connect a CAT5 cable from LAN port of customer-supplied computer to second port of
Ethernet switch.
Figure A-6. HMI Server Connection Ports
OPERATING HMI SERVER
The HMI Server System permits continuous monitoring of up to four centrifuges from a single
computer located in a non-hazardous area. A combination of DE-1000 and DE-7200 centrifuges
may be monitored. All information displayed on the local Operation screen of each centrifuge may
be viewed on the computer. System flexibility allows the operator to view all machines
simultaneously or individually.
Startup and Shutdown
Both the HMI Server and Router Switchgear Box require 220/240Vac 50/60Hz power for
operation. Units are turned on separately. Proceed as follows to turn on equipment:
1. Apply power to the HMI Server. PC will start up once power is applied. If PC does not start up,
press power button or cycle power to PC.
2. Once HMI Server boots up, the Home screen is displayed.
3. To shut down HMI Server, simply press and release power button on Server. If Server is
inaccessible, select Exit Runtime from the Setup screen. Next, select Shutdown PC on the
DeskLock Launcher.
Note! If Server is shut down, restart by cycling power or pressing power
button. Wait at least 30 seconds before re-applying power to restart the unit.
4. Turn off power to Router Switchgear Box to shut down unit.
Operation
If an existing network infrastructure having a DHCP server is not available to connect to LAN 2 of
the HMI Server, a four-port router with DHCP capability may be used. Alternatively, the LAN 1
interface may be used on the HMI Server PC to connect to the HMI Server. If desired to connect
to the LAN 1 interface, a basic 10/100 Mbps Ethernet switch will be needed. The switch would be
installed between the router box and the HMI Server PC (instead of a direct connection).
Additional PCs would then connect to this new customer-supplied switch.